Labconco Canopy Connections for XPert Filtered Systems and TXE Enclosures

• Allows enclosure systems to be thimble-ducted to the outside for additional protection
• The canopy's air gap provides a means of balancing air by allowing room air to enter the ductwork and mix with the filtered air being exhausted from the system
• House exhaust or remote blower is required (not included)†

†When sizing a blower for thimble-ducted systems, add 25% to the total air volume exhausted to allow for air entering around the canopy gap

Image(s) are representative of the product group and not necessarily the individual product.
Read More
• Allows enclosure systems to be thimble-ducted to the outside for additional protection
• The canopy's air gap provides a means of balancing air by allowing room air to enter the ductwork and mix with the filtered air being exhausted from the system
• House exhaust or remote blower is required (not included)†

†When sizing a blower for thimble-ducted systems, add 25% to the total air volume exhausted to allow for air entering around the canopy gap

Image(s) are representative of the product group and not necessarily the individual product.